The Cetacean Research Technology C304

Linear Frequency Range (±3dB) [kHz]: 0.007 to 250+
Usable Frequency Range (+3/-20dB) [kHz]: 0.005 to 250+
Transducer Sensitivity* [dB, re 1V/µPa]: -203
Preamplifier Gain [db]: 34
Effective Sensitivity* [dB, re 1V/µPa]: -169
SPL Equiv. Self Noise at 1kHz [dB, re 1µPa]: ~ 63
Power Requirement [Vdc]: 5 to 32
RMS Overload Acoustic Pressure [dB, re 1µPa]: 173 to 190
Maximum Operating Depth ** [m]: 460
Operating Temperature Range [ºC]: -40 to 60
Output Impedance [
Dimensions [mm]: 83 L x 32 W x (25 to 6)T
Integral Connector: Subconn MCBH3MSS
* Transducer Sensitivity + Preamplifier Gain = Effective Sensitivity.
** 1,000 meter versions are available. Append "-1k" to the model number to order a hydrophone with a depth rating >1,000 meters.
Shielded Cable Length [m]
• mating connector molded to 15, 30, or 50 [m] shielded cable
• other lengths available on a custom order basis
• cable includes Integral Connector rated to a depth of 7,000 meters.
Battery / Connector box
Battery boxes are rated NEMA-4X water resistant. Standard signal connection for the Cetacean Research™ C304 hydrophone is through a watertight female BNC connector. The hydrophones are powered by a single 9V battery (not included).
Signal Output: female BNC
Battery Life [hours]: > 20
Box Dimensions [mm]: 98 long x 64 wide x 40 high
Directionality
• omnidirectional below 10kHz;
• very directional at high frequencies (e.g., ~15º solid angle at 200 kHz, -3dB);
• directionality pattern is a dipole.
Options
• removable in-line high-pass filter
• external power supply support
• custom orders not listed above
C304 frequency response
C304 beam pattern at 200kHz
[Sleeker_special_clear]